An
insight
One observant
instrument user has noticed that our EGT/CHT readings tend to
be higher than with ordinary gauges - Why is that ?
Our EGT/CHT
gauges, like all others are based on thermocouples. Instruments
based on these devices can only read correctly if the temperature
of the "cold junction" (i.e. the terminals where the thermocouple
connects to the instrument) is known. The instrument needs to
measure this temperature and it needs to be taken into account
in the overall calculation of the temperature at the tip of the
thermocouple.
It may surprise
you, but very, very few instruments do this. It is common practice
to adjust the instrument during manufacture so it reads correctly
with the instrument at zero degrees Celsius (the Ice point). This
unfortunately means there is an error if the temperature of the
instrument is not at zero degrees. For example, should you be
at a more manageable 25 degrees Celsius - your instrument will
UNDER READ by about 25 degrees !!! Our instruments have very high
measurement resolution, normally only found in laboratory instruments
and they compensate for the local temperature. This is called "Cold junction compensation". As a result our readings are very
accurate and this does not change with ambient temperature as
the instrument automatically adjusts itself.
A further,
unusual measure found in our instruments related to measuring
temperatures, is that the amplifier used to amplify the very small
voltages from the thermocouples is itself measured for errors
caused by ageing and drift - and these are corrected for in the
measurement result.